Imtakt

YAZAWA HPLC COLUMN SCHOOL

JAPANESE

Do You Know Your Gradient Mixer Volume?
Gradient Delay Time

When running gradient analysis, many people carefully check column and mobile phase conditions, but surprisingly few pay attention to the mixer volume of their instrument. However, when performing fast gradient analysis with short or narrow-bore columns, the mixer volume can have a major impact on the chromatogram.

■ What happens when the mixer volume changes?
The figure above shows five peptides analyzed under identical conditions, with only the mixer volume changed to 0.1 mL, 0.5 mL, and 1.7 mL.
As the mixer volume increases, the analysis time becomes longer, peak heights decrease, and even the separation pattern changes.

■ The cause: "delay time"
Even when the pump starts the gradient, the mixed mobile phase does not reach the column immediately. It takes time to pass through the mixer volume (the delay volume, or dwell volume), and this time is called the delay time.

Delay time = Delay volume / Flow rate

In the example above, the delay time is 0.1 mL ÷ 0.3 mL/min = 0.3 min, and 1.7 mL ÷ 0.3 mL/min = 5.7 min.

Because a volume of initial-composition mobile phase (12%B) equal to the mixer's internal volume (delay volume) continues to flow isocratically through the column, the sample undergoes unintended isocratic elution. Strongly retained components diffuse within the column during this time, resulting in broader peaks. On the other hand, weakly retained components elute before the gradient arrives, so they are hardly affected by the mixer volume. Because each component is affected differently, peak spacing and even the separation pattern change.

The shorter and narrower the column, the relatively greater the effect of the mixer volume. Even an instrument that caused no problems with a long 4.6 mm ID column may fail to reproduce results once you switch to a short 2 mm ID column.

■ Know your instrument's delay volume
The same method can give different results on different instruments. In method transfer or when comparing data with other laboratories, differences in delay volume between instruments are often the cause of discrepancies.

■ Choose a mixer suited to your column size
For fast analysis with short, narrow-bore columns, a small-volume mixer is advantageous.

■ Smaller is not always better
Reducing the mixer volume can lead to insufficient mixing, making baseline waviness and noise more likely. This is especially noticeable when detecting mobile phases containing TFA at low wavelengths (e.g., 220 nm), in which case a larger mixer volume is needed.

■ Differences between gradient systems
Low-pressure gradient systems mix the mobile phases with a single pump, so they require a large-volume mixer. High-pressure gradient systems, on the other hand, mix the mobile phases by controlling the delivery rates of two pumps, so mixing can be achieved with a small mixer.

(Reference) Differences in gradient system configuration

■ Summary
The reproducibility of gradient analysis depends not only on the column and mobile phase but also on the mixer volume of the instrument. It is important to know the mixer volume of the instrument you are using.


ZI27 / YAZAWA Itaru, hplc@imtakt.com