Imtakt

YAZAWA HPLC COLUMN SCHOOL

JAPANESE

Causes of variation in retention time and peak shape in HPLC analysis

There are several possible causes for changes in retention time and peak shape when switching HPLC columns.

1. Instrumentation environment

If both the column and the instrument were changed simultaneously, it is important to suspect changes in the instrument environment. In the case of a gradient system, differences in mixer capacity can lead to delays in the system, which then manifest as differences in retention time. To achieve consistent retention times in gradient analysis, such as for inter-room reproducibility, it is essential to use instruments from the exact same manufacturer, model, and even the same model version.

In low-pressure gradient systems, the mechanism for mixing the mobile phase differs between manufacturers. Therefore, when comparing instruments other than the same model, achieving consistent gradient slopes and corresponding retention reproducibility becomes quite challenging.

(Reference)
Differences in gradient systems

 

2. Column-to-column variation

Even when using the same instrument, sample, and mobile phase, replacing the column can result in different retention times. This is due to variations in the packing batch. Differences in ligand quantity and residual silanol content manifest as changes in retention and peak shape.

It is extremely risky to create a method based on a single column and then continue using that same column for actual analysis. Many issues arise when a column degrades after several years, and upon replacing it with a new column, reproducibility is lost.

It is impossible to manufacture columns with identical ligand molecule counts on the stationary phase surface. Once a method is created, it is essential to immediately verify the robustness of the method using a different column (preferably from a different packing batch) to ensure method validation.

 

3. Analytical conditions

When the analytical method is not robust, even between new columns, differences in retention, separation, and peak shape may occur. The primary cause of this is "insufficient ion strength of the mobile phase."

When analyzing ionic compounds with only water/organic solvents, there is a concern that the dissociation (high polarity) and non-dissociation (low polarity) states become incomplete, leading to instability in retention and peak shape. Even if pH adjusters are added to the mobile phase, insufficient concentration may prevent proper dissociation control of the solute, resulting in irreproducible results. For example, a 1mM salt concentration.

Silica-based stationary phases have acidic silanol groups on their surface, which tend to interact ionically with basic compounds, leading to retention. Additionally, the uneven distribution of silanol groups can cause peak shape distortion. When changing columns, the amount of silanol also changes, so in methods with insufficient ion strength, there is a risk of variation in peak retention and shape.

To ensure the same results when changing columns, it is necessary to increase the ion strength in the mobile phase. For example, using 50–100mM salt concentrations.

Some column manufacturers use pyridine/phenol as an example to demonstrate the benefits of endcapping, but even weak ions like pyridine can cause gradual peak changes during repeated analyses if the pH and ion strength in the mobile phase are not properly controlled. Moreover, if ionic impurities in the sample adsorb onto the stationary phase, they can have an ionic effect on the solute’s retention and peak shape. Analyzing strong ions, such as tertiary amines, with only water/organic solvents is even more risky.

If the solute structure contains ionic characteristics, it is essential to design the mobile phase with not only the appropriate pH but also adequate ion strength.

Even if the solute is not ionic, solute structures with polar groups, such as amides, can experience strong electrostatic interactions with silanol or siloxane groups. As a result, when changing columns, retention and peak shape may fluctuate. Increasing the ion strength of the mobile phase, as well as adjusting pH or temperature, can improve reproducibility.

It is crucial not to rely solely on a pH meter when adjusting the pH and ion strength of the mobile phase.

(Reference)
Method for preparing mobile phase buffer 

 

To improve inter-room reproducibility of the method, it is crucial to have a robust mobile phase design that ensures the same results even when changing columns.


YE28 / YAZAWA Itaru, hplc@imtakt.com