IMTAKT YAZAWA HPLC COLUMN SCHOOL

YAZAWA HPLC COLUMN SCHOOL

JAPANESE

Troubleshooting Issues with HPLC Pump Flow - Check Valve

This section explains troubleshooting when the mobile phase fails to flow normally from the HPLC pump.

While potential causes may include an empty mobile phase reservoir, pump flow rate setting errors, or pump system malfunctions, the most likely cause is a check valve failure. The check valve, also called a "non-return valve," controls the flow direction of the mobile phase by turning the flow path ON and OFF, ensuring it flows in only one direction. In HPLC, it typically consists of a "ball valve" and a "valve seat" that the ball rests against.

Although "serial pumps" exist, parallel pumps are globally the standard. In parallel pumps, two pump heads operate alternately, canceling out pulsations caused by the suction and discharge of the plunger.

Each of the two pump heads in a parallel double-plunger pump has an IN-side and OUT-side check valve. During normal flow, the ball valve lifts off the valve seat, allowing the mobile phase to pass through. Conversely, when the mobile phase attempts to flow backward, the ball valve seals tightly against the valve seat, preventing backflow. This mechanism ensures that the mobile phase flows consistently in one direction, even under high pressure.

Each pump head typically has two check valves (IN and OUT), and since most systems have two pump heads, the total number of check valves is four. Identifying which specific check valve is malfunctioning can be a highly challenging task.

The primary causes of check valve malfunctions in HPLC systems are as follows:

  1. Prolonged Inactivity:

    • When the HPLC system is left unused for an extended period, the check valve and valve seat may stick together and fail to separate, resulting in a complete stoppage of mobile phase flow.
  2. Debris Accumulation:

    • Debris lodged between the check valve and the valve seat prevents proper sealing, causing backflow of the mobile phase.

If the flow rate is reduced, the issue may not only involve the check valve but also air bubbles trapped inside the cylinder of the pump head. The following methods can help resolve the problem:

  • Use the "Purge Button": Perform a high-speed purge of the mobile phase using the purge function.
  • Create a Sudden Pressure Release: Attach an old column or a restrictor tube, start flowing at a high rate, and, once pressure builds up, quickly release the purge valve to atmospheric pressure. The sudden pressure difference can expel trapped air.

If there is an issue with the check valve, the following steps can help determine which one is not functioning properly:

  1. Remove Both OUT-Side Check Valves:

    • Detach the OUT-side check valves from both pump heads.
  2. Turn the Pump ON:

    • Check if the mobile phase flows out from the OUT-side openings.
  3. No Flow from the OUT-Side Opening:

    • If no mobile phase comes out from one of the OUT-side openings, the IN-side check valve on that side is faulty. Clean the corresponding IN-side check valve.
  4. Flow from the OUT-Side Opening:

    • If mobile phase flows out from the opening, the corresponding OUT-side check valve may be faulty. Clean the corresponding OUT-side check valve.

The following steps outline recommended methods for cleaning check valves, distinguishing between IN-side and OUT-side valves:

  1. Initial Cleaning:

    • Place the problematic check valve (IN or OUT) into a beaker filled with distilled water.
    • Subject the valve to ultrasonic cleaning for approximately 10 minutes.
  2. Alternative Solvent Cleaning:

    • If the issue persists, replace the distilled water with ethanol or 2-propanol (IPA) and repeat the ultrasonic cleaning process.
  3. Acid Cleaning:

    • If the problem remains unresolved, fill the beaker with approximately 1M nitric acid solution (take precautions to avoid skin contact) and perform ultrasonic cleaning.
    • Afterward, thoroughly clean the valve again using distilled water in the ultrasonic cleaner.
  4. Manual Disassembly and Cleaning:

    • If the above steps do not resolve the issue, carefully disassemble the check valve unit and clean the ball valve and valve seat. Be cautious, as reassembly may not always be possible.
    • If manual cleaning is not feasible or successful, replace the check valve with a new one.

Maintaining the performance of high-pressure pumps in HPLC is not an easy task. While having maintenance performed by the manufacturer under a service contract is a valid approach, chromatographers should acquire the basic skills to perform simple maintenance tasks, such as cleaning and maintaining check valves, on their own.


YA10 / YAZAWA Itaru and TACHIKAWA Hiroshi, hplc@imtakt.com]

| HOME | PRODUCTS | NEWS | SUPPORT | OTHERS |